Accurate: ±0.1% of reading from 10% to 100% of the transducer
measurement range.
Repeatable:
±0.05% of reading or ±0.2 in-oz, whichever is greater,
based on an average of five readings.
Readings
displayed in ft-lb, in-lb, in-oz, n-m, dn-m, m-kg, and cm-kg units.
"Track,"
"Max," and "First Peak" operating modes available.
Worm
gear torque loading mechanism permits precision control of the applied
load for maximum accuracy and repeatability.
Electromagnetic
brake on torque loading mechanism prevents damage to test items
due to over-torquing.
Microprocessor
control prompts the operator through the setup and calibration process.
Removable
Lexan polycarbonate safety shield.
RS-232
serial interface permits data logging using a PC compatible computer.
Dead
weight recalibration kit (P/N 3714-15-1) available for field maintenance.
Portable:
Torque calibrator unit, control console, electrical cables, and
accessories can be packed in a rugged polymer storage case (included)
for convenient transport from site to site.
Optional
rechargeable battery pack with built-in AC charger permits torque
calibrator to be operated in areas where other power sources are
unavailable.
|